To set rates accurately in doserate or flowrate increments it may be necessary to switch between
the rate adjust options SET BY DOSERATE and SET BY ml/h. An arrow to the left of the rate
display indicates the rate that will change when the chevron buttons are used to
increase/decrease the infusion rate.
To set a doserate precisely the arrow must be pointing to the doserate (mg/kg/h); the flowrate will
be calculated from the doserate.
To accurately set a flowrate the arrow must be pointing to flowrate (ml/h); the doserate will be
calculated from the flowrate.
7001FAOPT73 ISS 13
Set by Doserate / Set by ml/h
Selecting the Set By Doserate Option
1. Press the ? button to access the options
menu.
2. Select the SET BY DOSERATE option
using the chevron keys.
3. Press the OK softkey. This will select the
set by doserate option and an arrow will
automatically point to the doserate on the
display. If necessary the doserate can be
increased or decreased using the chevron
keys.
Selecting the Set By ml/h Option
1. Press the ? button to access the options
menu.
2. Select the SET BY ml/h option using the
chevron keys.
3. Press the OK softkey. This will select the
set by flowrate option and an arrow will
automatically point to the flowrate on the
display. If necessary the flowrate can be
increased or decreased using the chevron
keys.
Getting Started
13 / 87