0% Uт (100% dip
0% Uт (100% dip
Voltage dips
in Uт) for 0.5 cycle
in Uт) for 0.5 cycle
IEC 61000-4-
at 50 Hz or 1 cycle
at 50 Hz or 1 cycle
11
at 60 Hz
at 60 Hz
70% Uт (30%
70% Uт (30%
Short
dip in Uт) for 20
dip in Uт) for 20
interruptions
cycles at 50 Hz or
cycles at 50 Hz or
61000-4-11
30 cycle at 60 Hz
30 cycle at 60 Hz
Voltage
0% Uт (100%
0% Uт (100%
variations on
dip in Uт) for 250
dip in Uт) for 250
power supply
cycles at 50 Hz or
cycles at 50 Hz or
input lines
300 cycle at 60 Hz
300 cycle at 60 Hz
61000-4-11
Power
frequency
magnetic
30 A/m
30 A/m
fields
(50/60Hz) IEC
61000-4-8
NOTE : UT is the main voltage (AC) prior to the application of the test level.
▪
Guidance 2
... ...
... R ecommended separation distances between portable and
mobile communication equipment and the i600. The i600 is
intended for use in an electromagnetic environment in which
Mains power quality
radiated RF disturbances are controlled. The customer or the
should be that of a
user of the i600 can help prevent electromagnetic interference
typical commercial
or hospital
by maintaining a minimum distance between portable and
environment. If the
mobile RF communications equipment (transmitters) and the
user of the i600
i600 as recommended below, according to the maximum
image intensifier
requires continued
output power of the communications equipment.
operation during
power mains
Rated
interruptions, it
maximum
is recommended
output
that the i600
power
image intensifier
of the
be powered from
transmitter
an uninterruptible
[W]
power supply or a
battery.
0.01
Power frequency
0.1
magnetic fields
1
should be at levels
characteristic of
10
a location in a
100
typical commercial
For transmitters rated at a maximum output power not listed above, the recommended
or hospital
separation distance d in meters (m) can be estimated using the equation applicable to
environment.
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1...
NOTE 2...
Separation distance according to the frequency of
transmitter [m]
IEC 60601‐1‐2: 2007
150 kHz to
80 MHz to
800 MHz to
80 MHz
800 MHz
2.5 GHz
d = 1.2√P
d = 1.2√P
d = 2.3√P
0.12
0.12
0.23
0.38
0.38
0.73
1.2
1.2
2.3
3.8
3.8
7.3
12
12
23
... A t 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
... T hese guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
Intraoral Scanner i600
IEC 60601‐1‐2: 2014
150 kHz to
80 MHz to
80 MHz
2.7 GHz
d = 1.2√P
d = 2.0√P
0.12
0.20
0.38
0.63
1.2
2.0
3.8
6.3
12
20
19