7.
Save the data by clicking "Save".
8.
Remove the SmartPeg by using the SmartPeg Mount.
9.
The instrument is turned off by pressing the on/off key for 2 seconds
10.
To remove the probe from the instrument, pull gently straight out, holding around the probe connector.
Do not pull the cable and do not twist the connector.

12) View stored Measurements

Enter the patient list
, choose patient and click on the implant to view.
A blue circle indicates the selected implant and the tooth number is
shown above. All measured data is found on the display. The first
measured ISQ value is shown to the left. To show the remaining ISQ values,
slide your finger from right to left. Below each ISQ value, the date of meas-
urement is shown. The values can be graphed by clicking

13) How to Measure on an Abutment

When a measurement is made on an abutment or on an implant with a "built-in" abutment, the ISQ value will be
lower compared to a measurement made on the implant. This is due to the difference in height above bone level.
To find out the ISQ difference to the measurement performed at implant level, a measurement should be taken
on the implant before the abutment is attached and then a second measurement on the abutment.

14) How to Measure in Classic Mode

To choose Classic mode, see section 8) Preparation and Set-up, Exam mode section.
1. Connect the Probe to the instrument.
2. Start the instrument
3. Place the SmartPeg into the SmartPeg Mount
SmartPeg Mount will hold the SmartPeg. Attach the SmartPeg to the implant or abutment
by screwing the SmartPeg Mount using finger force of approximately 4-6 Ncm
Do not over-tighten, to avoid damaging the SmartPeg threads.
4. Hold the probe close (3-5 mm) to the top of the SmartPeg without touching it
in both the Buccal-Lingual
An audible sound indicates that a measurement has been performed and the value, time and
date are shown on the display. Measured values can be deleted by clicking the trash can.
Note that values cannot be deleted after saving.
5. Remove the SmartPeg using the SmartPeg Mount.
6. The instrument is turned off by pressing the on/off key for 2 seconds
7. To remove the probe from the instrument, pull gently straight out, holding around the probe connector.
Do not pull the cable and do not twist the connector.

15) View Measurement in Classic mode

The ISQ values are shown directly on the screen. To browse between the measured values, use the arrows.
32 32
.
1
4
and Mesial-Distal direction
.
.
. The SmartPeg is magnetic, and the
5
. Do not measure from above
.
1
2
2
.
3
. Measure
6
.
4-6 Ncm
25070-05 ML