Instructions for Use
Angulus2 Angled Screwdriver
Loosen the head from the sheath by turning it from
•
Remove the sheath from the head.
•
Rinse under demineralized water < 35 °C / < 95 °F (1).
•
•
Remove any liquid residues with absorbent, lint-free cloth. Then blow dry with
compressed air (2).
NOTICE
Risk of material damage!
Do not place the angled screwdriver in disinfectant solution or in an ultrasonic bath.
•