Instructions for Use
Angulus2 Angled Screwdriver
Ensure that the angled screwdriver is completely dry, both inside and out, following thermo
•
washer disinfection.
Remove any liquid residues using compressed air (maximum operating pressure of 3 bar).
•
•
Oil the dry angled screwdriver immediately after thermodisinfection.
6.2.5
Reassembling the Angled Screwdriver
Reassemble the angled screwdriver after manual and mechanical cleaning and disinfection.
•
NOTICE
Risk of material damage!
All 3 EN (serial numbers) must match. Possible damage to products in the case of a failure to
comply.
•
Insert the head into the sheath until it engages.
Pay attention to the symbols and turn from
•
Revision 04
to
until it locks.