Instructions for Use
Angulus2 Angled Screwdriver
Danger of infection due to non-sterile handling!
Improper sterilization and non-sterile handling of the instrument can lead to serious health
hazards for patients.
Sterilization must be carried out according to a validated steam sterilization process, for
example in a sterilizer complying with EN 13060 Class B/EN 285.
ANSI/AAMI ST79: Sterilization temperature 134 °C (273.2 °F) with a minimum holding time
of 3 min (potential prion contamination: 18 min). Please follow the instructions of the user
manual of your steam sterilizer.
Due to the different packaging materials, ambient conditions, quality of the steam as well as
the materials and the total weight of the products which have to be sterilized, the performance
features of the sterilizer and the variable cooling times the duration of drying times may differ
substantially. The effective necessary drying time is therefore determined by the user.
Make sure that you only remove dry sterile goods from the sterilizer.
Revision 04