be increased or decreased quickly by
pressing and holding the ▲ ▼ buttons.
Mode
Press the mode selection button to switch
between modes (TENS, EMS I, EMS II, EMS III
or Massage).
Programme
Press the programme to change the
selected programme.
Settings
Press the settings button to change the
settings of selected programmes. See
full directions for changing settings in
section 5.
Phase Skip
In EMS programmes, press phase skip
button to skip to the next phase of the
programme.
Lock / Unlock
If no key is pressed for 30 seconds, Unipro
will auto-lock. To unlock the keypad, press
either CH1- or CH2-.
Keypad Lock Icon
When the keypad is unlocked, the icon will
be grey. When the keypad is locked, the
icon will be coloured. If the icon is flashing,
this means a button has been pressed
whilst the device is locked.
Warning Icon
The warning icon will flash in EMS
manual
programmes
period is less than Rest Time=Work
Time*(WorkHz-16.66)/16.66 seconds.
if
the
Rest
Connection Error Icon
If the device is incorrectly connected to
the body, this the connection error icon
will flash.
05.
Directions for
Use
Connecting the Lead Wires
and Pads
1. Insert the lead wire plug into the base
of the unit.
2. Attaching the electrode pads to the
lead wire. Push the pin ends firmly into
the pigtail ends of the electrode pads.
3. Before applying the electrode pads to
your body, make sure your skin is clean
and dry.
4. Place the electrode pads on the skin
(see Pad Placement).
Warning: Ensure Unipro is switched
off before applying the electrode
pads on the skin.
Turning Unipro On
To turn the unit on, press the
button on the side of the device for
approximately 1 second until the display
shows. The unit will start with the
programme you used last.
Unipro will automatically turn off if both
strength controls are set to zero for 5
minutes.
Power