15.2 Autoclave sterilization (MulTipeg & MulTipeg Driver)
Sterilization should be madein a pre-vacuum steam sterilizer (autoclave) according to ISO 17665-1.
Clean the products and put them in an FDA-cleared (USA) autoclave bag before sterilization.
The following sterilization process shall be used:
At least 3 minutes at 134 (-1/+4)°C or 273(-1.6/+7.4)°F
Follow the instruction for the autoclave that isused.
Do not clean the MulTipeg by ultrasound. Could cause damage
The instrument must be used with a cover in all uses. (Only US)
The instrument must be cleaned with a disinfectant between patients.
15.3 Cleaning
Instrument
The instrument is cleaned with a wet cloth. Any of the above disinfectants can be used. For use in environments requiring sterility, the instrument should be
covered with a sterile cover. Recommended covers (US) are Probe Covers, for instance #DYNJE5900 by Medline or part #10040 by Medical Resources.
MulTipeg Driver and MulTipegs
The MulTipeg Driver and the MulTipeg should be cleaned with water with or withoutdetergent, using a light brush. Any of the above disinfectant fluids can be
used. For usein environments requiring sterility, the MulTipeg Driver and MulTipeg should be autoclaved before use according to the above instructions.
16. Lifetime
The lifetime of the instrument is determined by the number of charging cycles. The internal batteries can be fully charged more than 500 times before they need to
be replaced. The instrument should not be left uncharged for more than 1 year.
The MulTipeg Driver is guaranteed for at least 100 autoclave cycles, and a MulTipeg is guaranteed for at least 20 autoclave cycles, before they aredegraded in any
way.
17. Troubleshooting
17.1 Possible errors
Difficult to achieve ameasurement:
In some cases it is more difficult for theinstrument to bring the MulTipeg into vibration. If so, try to hold theinstrument tip closer to thetip of the MulTipeg.
Check also that no soft-tissue is touching the peg which might stop its vibration.
Noise warning (audible and visible on thedisplay):
An electric device close to the instrument is causing the warning. Try to remove the source.
The instrument suddenly turnsoff:
Theinstrument turns off automatically after 30 seconds of inactivity. It also turns off if the battery level is too low, and due to any of the error codes described
below.
Not all segments are lit up when instrument is started:
The instrument is damaged and has to be sent for repair or exchange.
17.2 Error codes
If malfunctioning, these error codes are shown on the display before it turns off:
E1: Hardware error. Malfunctioning electronics
E2: Noise error. Shown if constant electromagnetic noiseis present
E3: Pulse power error. Malfunctioning magnetic pulse generation
Useof accessories other than thosespecified or provided by the manufacturer of this equipment could result
in increased emissions or decreased electromagnetic immunity of this equipment and result in improper
operation.
18. Accessories & Spare Parts
Model
MulTipeg Driver
REF
55003
MulTipeg: Please refer to the updated list from the supplier.
19. Pairable instruments
Product
NSK Surgic Pro2 (Control unit)
20. Service
In case of a malfunctioning instrument, thedevice should besent to the manufacturer or distributor for repair or exchange.
Osseo 100+ is covered by a two-year warranty.
Sterile Cover
Mains adapter
55105
55093
incl 1.5 m cable
Fuhua Electronic Co., Ltd
UE05WCP-052080SPC
Model
NE335
EU plug
UK plug
AU plug
55094
55095
55096
US plug
55097