WARNINGS
Please read carefully the instructions given below, as they supply you with all the needed information, necessary for
installation, use and maintenance.
Once you receive the V-MS check out for its integrity and all of its components, in case of any anomalies, please consult
a skilled staff before making any operation. The equipment has to be operated by skilled persons.
For a correct functioning it is necessary to use original spare parts and original accessories. Microdos srl declines whatever
responsibility in reference to break down due to tampering or the use of not original spare parts and accessories.
The electrical plant has to be in conformity with the rules of the country where it is realised.
During maintenance and repair of parts in contact with chemicals, always use personal protection measures (gloves,
aprons, goggles, etc.).
The usage room temperature can't over take 45 °C. The minimum temperature depends on the chemical that must remain
in the liquid state.
Microdos srl declines all responsibility in reference to any intervention on the equipment from a non skilled staff.
HAZARDS
Ignoring the instructions can result in equipment damage and, in extreme cases, injury to persons.
Before installing the equipment make sure that the electrical data reported on the label correspond to those of your
electrical plant.
All maintenance or repair must be performed with the system electrically isolated.
Whenever you need to access the connection box, turn off V-MS and isolate it electrically.
Do not operate on the equipment with wet hands or feet
Do not leave the equipment exposed to the action of atmospheric agents.
In case of an improper functioning of the equipment switch off and contact our technical assistance for any reparation
request.
WARNING : PROPER AND IMPROPER USE
The instrument V-MS should be used only for the purposes for which it was designed that is, the measurement and
control of the values of the chemical and physical parameters.
Uses other than these are considered improper and therefore dangerous.
The instrument is NOT designed for use in environment with danger of explosion.
Microdos srl declines any responsibility for any damage resulting from the improper and / or unreasonable use of the
instrument.
TECHNICAL FEATURES
All units are electrically isolated from the rest
Measure
pH (automatic compensation with temperature)
Orp (RX)
CL by CG_CL3
CL by CP_CLO_ M
CL by CP_CLI_M
CL by CP_CLI_B
CD by K1 probe
T by PT100 probe
Rev 6.0
Range
0.00-14.00 pH
Resolution: ±0.01pH
Precision: better than 0.1% full scale
0- 1000mV
Risolution: 1mV
Precision: better than 0.1% full scale
0.00-5.00 ppm
Resolution: 0.01ppm
Precision: better than 0.1% full scale
1.0 – 10.00ppm
Resolution: 0.01ppm
Precision: better than 0.1% full scale
0.00-2.00ppm
Resolution: 0.01ppm
Precision: better than 0.1% full scale
0µS – 20.00mS;
Resolution: 1µS sotto a 9999µS
Precision: 1% full scale
0.0°C- 99.9°C
Resolution: 0.1°C
Precision: 2% full scale
Cod. 00.008.764
0.01mS sopra 10.00mS