CONDUCTIVITY PROBE CALIBRATION
Connect the conductivity probe to pins 22, 23 and 24 of the board as shown below.
For the wiring of the probe code 00.043.618 refer to the labels on the cables (the cables of the PT100 must be connected
to connectors 15 and 16, taking care to short-circuit inputs 14-15 and 16-17)
OFFSET CALIBRATION
The Offset calibration is made in factory, it is therefore not necessary to do it.
If you need to change the sensor reading in the air follow the next instructions:
After connecting the conductivity probe to the connectors 22, 23 and 24 dry the probe and put it in the air. If the meter
reading is not 1uS, proceed to the Offset calibration.
Press Enter for 3 seconds to enter the
Navigation Menu and then move the cursor on
Calibration.
Press Enter to proceed with Calibration.
Move the cursor on Calibration CD and press
Enter.
Rev 6.0