RX PROBE CALIBRATION
Before starting the calibration, take care to prepare the buffer solution for RX calibration (any value) and dip the RX
(ORP) probe into the buffer solution.
Keep pressing Enter
button for 3 seconds to
enter in the main Menu.
Move the cursor to
Calibration and press
Enter to do the Calibration.
Move the cursor to
Calibration RX and press
Enter.
 
Edit (using
buttons) the mV reading
value of the buffer solution and press Enter.
After the calibration V-MS determines the
Quality of the Offset of the RX probe:
100% = Excellent probe
50% = Good probe to be monitored
25% = probe that you should change
0% = fully exhausted probe
While viewing the quality of the probe press Enter to return to the calibration menu or ESC to return to measurement.