3.5 Calibration
A calibration is performed similarly to the sample analysis described
before in chapter 3.4 using the Alivion Methanol Standard 1 and 2
WITHOUT/WITH ETHANOL.
A measurement is always evaluated with the most recent saved
calibration.
Regular calibration is essential to ensure maximum accuracy.
If a new calibration should be performed, the warning
is displayed on the home screen (see chapter 3.7 for details).
To perform a calibration, select Calibrate from the main menu followed
by the calibration mode:
Calibration without ethanol
Gives the maximum measurement accuracy when
measuring samples that do not contain ethanol.
Calibration with ethanol
Gives the maximum measurement accuracy when
measuring samples that contain ethanol.
NOTICE
Alivion recommends performing a calibration in the
beginning of each measurement day to achieve
maximum accuracy.
Alivion Spark M-30
Measurement with Alivion Methanol Standard 1
WITHOUT ETHANOL (orange liquid, red bottle cap)
Measurement with Alivion Methanol Standard 2
WITHOUT ETHANOL (blue liquid, red bottle cap)
Measurement with Alivion Methanol Standard 1 WITH
ETHANOL (orange liquid, blue bottle cap)
Measurement with Alivion Methanol Standard 2 WITH
ETHANOL (blue liquid, blue bottle cap)
Use
13