Troubleshooting
7 Troubleshooting
This section explains device faults and provides instructions on how to
resolve them.
7.1 Sampling errors
Timeout during sampling
Remove the sample vial within 10 seconds after the Alivion
Spark M-30 displays Remove sample.
Make sure to confirm the removal of the sample vial within
10 seconds.
Sampling vial not connected correctly
Make sure an Alivion sample vial is screwed into the Alivion
Spark M-30 before pressing Continue on the Insert sample
screen.
Make sure only original Alivion sample vials are used.
Make sure the vial is screwed in hand-tight.
Retry the measurement with a new sample vial.
Check if the sealing O-ring (see chapter 2.1) is missing. If so,
replace it with the included replacement or otherwise contact
Alivion's technical support.
Check if the windings are damaged in any way. If so, please
contact Alivion's technical support.
Contact the Alivion technical support if the problem persists
Sampling failed (ERR 942)
Do not remove the sample vial before prompted to do so.
Ensure that the sample vial is not moved or tilted during
sampling.
Make sure the vial is screwed firmly and try using a new vial.
Make sure the sealing O-Ring is installed correctly.