7.4 General errors
Critical error (ERR 900)
Restart the device and retry the measurement.
Contact the Alivion technical support if the problem persists.
Charging not possible during measurements
Remove the charging cable from the device and retry your
measurement.
Disconnect the Alivion Spark M-30 from the computer and retry
your measurement.
Otherwise restart your Alivion Spark M-30 without a cable
connection.
Please clear inlet
Make sure nothing blocks the inlet of the Alivion Spark M-30.
Remove any sample vials or debris at the inlet.
Contact the Alivion technical support if the problem persists
Temperature not within operational range
The device can only be used for measurements within the
operational temperature range of 10–35 °C.
Make sure the ambient temperature is within the operational
temperature range, let the device equilibrate and retry the
measurement.
Measurement series is full
A maximum of 100 measurements can be saved in the active
data series.
Create a new data series under:
Settings → Resets → New measurement series.
My results seem not accurate/consistent
Ensure there are no warnings are active (see chapter 3.7).