Contact Alivion's technical support if the problem persists.
Sampling failed (ERR 943)
Make sure only Alivion vials are used and try using a new vial.
Contact Alivion's technical support if the problem persists.
7.2 Measurement errors
Measurement failed (ERR 950/951)
Ensure the inlet of the Alivion Spark M-30 is clean. Remove
debris or liquid sample at the inlet of the device.
Do not screw in a sample vial or block the inlet in any other way
during the analysis.
Contact Alivion's technical support if the problem persists.
Measurement failed (ERR 960/961)
Retry the measurement with a freshly prepared sample.
Retry the measurement with a different sample liquid.
Try to clean the sensor (Settings → Advanced → Clean sensor)
and re-calibrate the device.
Contact Alivion's technical support if the problem persists
Critical error (952/953)
Restart the device and retry the measurement.
Contact Alivion's technical support if the problem persists.
Ethanol concentration too high (954)
The Spark M-30 is only able to measure samples with an
ethanol concentration up to 20 vol%.
Contact Alivion's technical support if the problem persists.