Troubleshooting
7.3 Calibration errors
Calibration failed. Measurement error
The calibration failed due to a measurement error described in
chapter 7.2. Follow the instructions for the respecting
measurement error.
Temperature difference during calibration too large
Ensure the ambient temperature does not change during the
first and second calibration measurement. Retry the calibration
once the temperature is constant.
Calibration failed (ERR 972 -976)
Ensure the correct standard was used for the calibration.
For calibration without ethanol Alivion Methanol Standard 1
WITHOUT ETHANOL (orange liquid, red bottle cap) is used first,
followed by Alivion Methanol Standard 2 WITHOUT ETHANOL (blue
liquid, red bottle cap).
For a calibration with ethanol Alivion Methanol Standard 1 WITH
ETHANOL (orange liquid, blue bottle cap) is used first, followed
by Alivion Methanol Standard 2 WITH ETHANOL (blue liquid, blue
bottle cap).
Ensure only original Alivion calibration standards and sample
vials are used.
Ensure the inlet of the device is clean and free of debris or
liquid sample.
Ensure any used sample vials and pipettes are not near the
inlet of the Spark M-30.
Ensure no sample liquid is on the windings or rim of the sample
vials.
Try to clean the sensor (Settings → Advanced → Clean sensor),
reset the calibration to factory settings (Settings → Resets →
Reset calibration) and retry to calibrate the device.
Contact the Alivion technical support if the problem persists.