12. Technical data
12.1 Classification according to Regulation (EU)
2017/745
Art. no.
Designation
GA330
Drilling and milling machine
12.2 Performance data, information about standards
Max. power
Max. speed
Direction of rotation
Cannulation
Weight (operational)
Dimensions (L x W x H, ready for
operation)
Application part
EMC
Conforming to standard
After 500 processing cycles, the manufacturer tested the product and it
passed.
12.3 Nominal operating mode
Operation with non-periodic load and speed changes (type S9 pursuant to
IEC EN 60034-1)
Drilling (clockwise/counterclockwise rotation)
■
60 second application, 60 second pause
■
20 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
Milling (clockwise/counterclockwise rotation)
■
30 second application, 30 second pause
■
8 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
Drilling (oscillation)
■
60 second application, 60 second pause
■
4 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
Saw operation using GB660R
■
30 second application, 60 second pause
■
3 repetitions
■
30 min cooling time
■
Max. Temperature 48 °C
Class
IIa
approx. 250 W
26 000 rpm
Right and left rotation, oscillation
4 mm
1.25 kg ±10 %
207 mm x 121 mm x 55 mm ±5 %
Type BF
IEC/DIN EN 60601-1-2
IEC/EN 60601-1
12.4 Ambient conditions
Operation
Temperature
10 °C to 27 °C
Relative
30 % to 75 %
humidity
Atmospheric
700 hPa to 1 060 hPa
pressure
13. Disposal
WARNING
Risk of infection from contaminated products!
►
Observe national regulations when disposing of or recycling the
product, its components, and their packagings.
Note
The operator must process the product before disposal, see Validated repro-
cessing procedure.
The recycling pass can be downloaded from the Extranet
as a PDF document under the respective article number.
(The recycling pass includes disassembling instructions for
the product, as well as information for proper disposal of
components harmful to the environment.)
Products carrying this symbol are subject to separate col-
lection of electrical and electronic devices. Within the
European Union, disposal is taken care of by the manufac-
turer as a free-of-charge service.
►
For questions regarding the disposal of the product, please contact your
local B. Braun/Aesculap representative, see Technical Service.
en
Transport and storage
-10 °C to 50 °C
10 % to 90 %
500 hPa to 1 060 hPa
13