14.2 Operating states/error signals

The product indicates operating states and error messages with beep signals.

14.2.1 Signals for operating states

Mode switching
Beep signal
1x short
1x long
2x short
3x short
3x short
4x short

14.2.2 Warnings/error signals

Error during use
Beep signal
3 x long
5 x long
10x long, then switch­
ing to empty battery
mode
Beep signal
1 x short at intervals
of approx. 1 second
Ongoing short beep
signal
Ongoing short beep
signal
Additional action performed
Take weight off leg
Button 1 or 2 was pressed long
or
Button 2 was pressed 2x short
Take weight off leg
Take weight off leg
Button 2 was pressed 2x short
Take weight off leg
Event
Battery charge level under 10 %
Battery charge level under 5 %
Battery charge level approx. 1 %
Event
Overheated hydraulic unit
Error of moderate severity:
e.g. a sensor is not operational
Switch to safety mode
Walking possible with restrictions
Loss of spring contact/loss of
sensor screw contact
Event
Switching to basic mode per­
formed successfully
When the beep signal sounds,
weight must be taken off the leg to
switch to the corresponding mode.
Switching to the additional mode
(2nd mode) performed successfully
Switching to LOM performed suc­
cessfully
Switching to LOM performed suc­
cessfully
No switching performed
Check the following points:
Was weight put on the leg with
the product or was there move­
ment in the knee joint?
Is the spring element in the foot
region
bent
defective?
Was calibration performed?
Required action
Reduce activity
Attempt to reset this error by con­
necting the battery charger. If this
error persists, the product must be
inspected by an authorised
Ottobock Service Centre.
The beep signal can be muted by
pressing the top button on the joint
unit (5 seconds).