E N G L I S H
K. ARMREST RECEIVER ATTACHMENT
Adjusting Armrest Receiver Fit
To tighten or loosen the fit of the outer armpost in the receiver:
1. Loosen the four receiver adjustment bolts (D) on the sides of the receiver.
2. With the armrest in the receiver (E), squeeze the receiver to achieve the desired fit.
3. Tighten the four bolts (D). (144 in-lbs, 16.3 Nm)
Position Adjustment
1. Loosen the two clamp bolts (10) until clamp is loose.
2. Slide armrest receiver to desired position.
3. Tighten both bolts to 62 in.-lbs. (7 Nm).