Dynamic Trial Fitting; Checking The Factory Settings; Setting Flexion Damping - Ottobock 3R106-Pro Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 61
+ = anterior placement / – = posterior placement (in relation to the load line)
Item
Statischer Aufbau
   
m–l positioning, centre of the prosthetic knee joint to the frontal load line
1 to 2 cm medial
m–l positioning of the prosthetic socket:
Positioning of the anterior superior iliac spine (Spina iliaca anterior superior)
line
:
1 to 2 cm lateral

5.5 Dynamic Trial Fitting

CAUTION
Adjusting the Settings
Fall due to incorrect or unfamiliar settings
Only adapt the settings to the patient gradually.
Explain the effects of the adjustments on the use of the prosthesis to the patient.
During dynamic trial fitting, the alignment and settings of the prosthesis are checked and adapted
for optimum walking according to the needs and abilities of the patient.
The patient has to learn the safe use of the prosthesis through intensive training.
The following subsections describe the adjustment possibilities for adapting the product to the
patient.
The list below provides an overview of the sequence for working through the subsections:

Checking the Factory Settings

Setting Flexion Damping

Setting Extension Damping
5.5.1 Checking the Factory Settings
Function
Flexion damping
Extension damping
Tools required:
>
710H10=2x3 Adjustment Wrench
1) Check whether the settings match the factory settings (see table).
2) In case of deviations, restore the factory settings using the adjustment wrench.
5.5.2 Setting Flexion Damping
Function
Flexion damping
Tools required:
>
710H10=2x3 Adjustment Wrench
► CAUTION! The prosthetic foot must not swing through too far.
Adjust flexion damping using the adjustment wrench so that the prosthetic foot swings
through sufficiently according to the needs of the patient.
Static alignment process
Means of adjustment
Adjustment valve "F"
(see fig. 4 -
)
Adjustment valve "E"
(see fig. 4 -
)
Means of adjustment
Adjustment valve "F"
(see fig. 4 -
)
Factory setting
From the left-hand stop,
turned to the right (num­
ber of turns: 2.0)
Turned all the way to the
left
Setting
Rotation to the right
Rotation to the left
:
to the frontal load
Meaning
Valve slightly closed –
low damping as the start­
ing point for adjustment
to the patient
Valve fully open – minimal
damping as the starting
point for adaptation to the
patient (safety setting)
Meaning
Increase damping
Reduce damping
29

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

3r106-pro-kd3r106-pro-st

Tabla de contenido