9.2
Repairs
For repairs, contact Customer Service Center in the USA at +1-800-382-4682 . For repairs
outside the United States, contact your local affiliate . Relevant service instructions are provided
in DePuy Synthes work instructions . DePuy Synthes will provide circuit diagrams, component
part lists, descriptions, calibration instructions, or other information that will assist service
personnel trained and authorized by the manufacturer in repairing the pump .
9.3
Electrical Safety Testing Information
For electrical safety testing information, contact Customer Service Center in the USA at
+1-800-382-4682 . Outside the United States, contact your local affiliate .
9.4
Fault Codes
The FMS VUE II software detects the following fault conditions .
Fault Code
FC001
System Error
(Unrecoverable)
FC002
System Error
(Recoverable)
FC003
Pressure tubing system problem
detected
Table 10. Fault Codes
Fault Description
56
Fault Correction
Discontinue use of the console . Contact
your DePuy Synthes representative .
Power down the pump .
Wait 1 minute before power up .
If fault occurs again, discontinue use of
the console .
Contact your DePuy Synthes
representative .
Pressure sensors readings are matching
but incorrect .
Change the Tube Set . If changing Tube
Set does not correct issue, discontinue
use of the console . Contact your DePuy
Synthes representative .