Johnson & Johnson DePuy Synthes FMS VUE II Manual Del Usuario página 72

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 73
Recommended Separation Distances between portable and mobile
RF Communications equipment and the FMS VUE II
ME Equipment and ME Systems that are NOT Life-supporting
Recommended Separations Distances for FMS VUE II and Shaver
The FMS VUE II is intended for use in the electromagnetic environment in which radiated
disturbances are controlled . The customer or user of the FMS VUE II can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF Communications Equipment and the FMS VUE II as recommended below, according to the
maximum output power of the communications equipment
Max Output
Separation (m)
Power
150kHz to 80MHz
(Watts)
D=(3.5/V1)( √ P)
0 .01
0 .11667
0 .1
0 .36894
1
1 .1667
10
3 .6894
100
11 .667
Frequency
RF Output Power (EIRP)
Modulation
Type of Antenna
Separation (m)
for ISM Bands
150kHz to 80MHz
D=(3.5/V2)( √P)
0 .05833
0 .18447
0 .58335
1 .8447
5 .8335
RFID Transmitter/Receiver Specifications
Field Strength = 60 .73 dBµV/m @ 10meters
ASK (Amplitude Shift Keying)
Spiral Loop PCB Antenna (Gain = 0 dBi)
72
Separation (m)
80 to 800MHz
D=(3.5/E1 ( √ P)
0 .11667
0 .36894
1 .1667
3 .6894
11 .667
13 .56 MHz
EIRP = -24 .04 dBm
Separation
(m) 800MHz to
2.5GHz
D=(7/E1)( √P)
0 .233333
0 .73785
2 .3333
7 .3785
23 .333

Publicidad

Tabla de contenido
loading