Tunneling the Lead or Lead Extension

Tunneling Tool Assemblies (SC-4254 and SC-4252) are sold separately. If using a 4x8 Surgical Lead or
Infinion CX Lead, we recommend using the 35 cm Long Tunneling Tool (SC-4254).
Note: The following Codman Disposable Catheter Passers may be used in place of the Boston Scientific
tunneling tool: REF 82-1515 (36 cm); REF 82-1516 (55 cm); REF 82-1517 (65 cm) When using a
Codman Disposable Catheter Passer, tunnel from the midline or anchor incision to the IPG pocket
using the standard technique.
1. If not already assembled, attach the Tunneling Tool handle to the shaft by turning the locking
mechanism clockwise.
Tool Handle
Locking Mechanism
2. Mark the desired route of the tunnel.
3. Administer the appropriate local anesthetic along the tunneling path
4. OPTIONAL. If necessary, bend the tool shaft to conform to the patient's body.
5. Make a small incision at the desired exit site.
6. Create a subcutaneous tunnel between the Lead(s) incision and the IPG pocket site until the straw is
visible and accessible at the exit point.
Shaft
Instructions for Use
Implantable Pulse Generator