Connecting The Lead, Lead Extension, Splitter, Or Connector To The 16 Contact Ipg; Connecting To The Ipg - Boston Scientific WaveWriter Alpha Instrucciones De Uso

Ocultar thumbs Ver también para WaveWriter Alpha:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 74

Connecting the Lead, Lead Extension, Splitter, or Connector to the 16 Contact IPG

Note: Lead Extensions, Adapters, and Splitters are not full-body MRI scan eligible.
IPG ports are labeled as follows:
C

Connecting to the IPG

• Leads connect to IPG ports C or D.
• For the Infinion 16, connect the Splitter tail with laser-etched bands (Contacts 1-8 of the Infinion 16
Lead) to the left Port C and the unmarked Splitter tail (Contacts 9-16 of the Infinion 16 Lead) to the
right Port D.
• For the Infinion CX Lead, connect the tail with the single marker band (Contacts 1-8 of the Infinion
CX Lead) to the Port C and the tail with two marker bands (Contacts 9-16 of the Infinion CX Lead
to the right Port D.
• For the Artisan and Artisan MRI 2x8 Surgical Leads, connect the left side to the left Port C. Connect
the right side (the laser-etched tail), Contacts 9-16, to the right port D.
D
Instructions for Use
Implantable Pulse Generator
92469342-02
9 of 353