reset
Reset [
]
Reset the Setup menu to factory defaults.
NO
= do not reset
yes
= reset
Capacity Unit [C.
Select the unit used for calibration.
t (Metric Tonne)
g
kg
lb
Note: t and g are not available for i-DT61XWE model.
[
]
rANGE
Range
Note: This setting is only avaliable for i-DT61XWE model.
Set the number of weighing intervals.
This terminals can be configured to use single or dual interval. Each interval can be assigned its own
graduation. If dual interval is selected, the graduation will change when the weight reaches the second
interval.
When Single interval is selected, the additional parameters available are:
>|1|< Capacity
>|1|< Graduation
When Dual interval is selected, the terminal functions with two intervals, each with its own capacity and
graduation. In addition to the Interval 1 capacity and graduation parameters, the following two parameters
are available:
>|2|< Capacity
>|2|< Graduation
[
] /
Cap
Capacity
Set the capacity of the scale or the capacity of the first scale (for i-DT61XWE model).
1...999999
[
] /
GrAd
Grad
Grad1
Set the scale readability or the readability of the first scale for i-DT61XWE model from 0.0001 to 100.
0.0001~100
[
Cap2
Capacity2
Set the capacity of the second scale for i-DT61XWE model.
[
]
GrAd2
Grad2
Set the readability of the second scale for i-DT61XWE model.
Power On Zero
Zero the scale at Power On.
Off
= disabled.
ON
= enabled.
[
Power On Unit
Set the unit that will be displayed at Power On.
AUtO
= last unit in use when turned off
kg
= kilograms
g
= grams
lb
= pounds
oz
= ounces
lb:oz
= pound ounces
t
= metric tonne (only available for i-DT61PW model)
UNIt
]
[
]
Cap1
Capacity1
[
]
GrAd1
]
[
]
P.ZerO
]
P.UNIt