Legal For Trade; Settings; Verification; Sealing - OHAUS Defender 6000 Manual De Instrucciones

Ocultar thumbs Ver también para Defender 6000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Defender 6000 Indicators
8.
If you need to check which COM (serial) ports your computer's Bluetooth is using, please follow the
following steps:
a) On the Bluetooth & other devices page, instead of clicking Add Bluetooth or other device, turn
down the page and find More Bluetooth options under Related settings.
b) Click More Bluetooth options.
c) Select COM Ports on the top bar, and you will see the using COM (serial) ports displayed.
Note: if you use other device to pair with the terminal, please follow the specific device's Bluetooth pairing
method.

5. LEGAL FOR TRADE

When the indicator is used in trade or a legally controlled application, it must be set up, verified and sealed in
accordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that
all pertinent legal requirements are met.
5.1

Settings

Before verification and sealing, perform the following steps:
1. Verify that the menu settings meet the local weights and measures regulations.
2. Perform a calibration as explained in Calibration Menu section.
3. Turn the indicator off.
To turn on the security switch:
1. Disconnect power from the indicator and open the housing as explained in Opening the Housing section in
the Installation chapter. (Remove the batteries for i-DT61PW model).
2. Set the position of the security switch SW1 to ON. (Check Mainboard section for where SW1 is located on
the mainboard).
3. Close the housing.
4. Reconnect power and turn the indicator on. (Re-install the batteries for i-DT61PW model).
5.2

Verification

The local weights and measures official or authorized service agent must perform the verification procedure.
5.3

Sealing

After the scale has been verified, it must be sealed to prevent undetected access to the legally controlled
settings. Refer to the illustrations below for sealing methods.
Figure 5-1. i-DT61PW Wire Sealing
Figure 5-2. i-DT61PW paper Sealing