Defender 6000 Indicators
4.4
Readout Menu
Enter this menu to set user preferences. Default settings are in bold.
Menu
Sub-Menu
Reset
Stability
Zero Range
Filter Level
Auto Zero Track
Bright Level (i-DT61XWE)
Back Light (i-DT61PW)
Read Out
(r.E.A.d)
Screen
Auto Off
P.SAVE (i-DT61PW)
L.KEY (i-DT61XWE)
End
[
]
reset
Reset
Reset the readout menu to factory defaults.
NO
= do not reset.
YES
= reset
[
StAblE
Stability
Set the amount the reading can vary before the stability symbol turns off.
0.5d
= 0.5 scale division
1d
= 1 scale division
2d
= 2 scale division
5d
= 5 scale division
[ZErO]
Zero
Set the percentage of scale capacity that may be zeroed.
2%
= zero range is +/-2%
100%
= zero range is +/-100%
[
]
FILtEr
Filter
Set the amount of signal filtering.
LOwW
= faster stabilization time with less stability.
mMEd
= normal stabilization time with normal stability.
HIGH
= slower stabilization time with more stability.
[
]
AZt
AZT
Set the automatic zero tracking functionality.
OFF
= disabled
0.5d
= the display will maintain zero until a change of 0.5 divisions per second has been exceeded.
1d
= the display will maintain zero until a change of 1 divisions per second has been exceeded.
3d
= the display will maintain zero until a change of 3 divisions per second has been exceeded.
Sub-Menu
(in
segment)
]
Options
reset
no, yes
Stable
0.5d, 1d, 2d, 5d
ZErO
2%, 100%
FIltEr
Low, Medium, High
AZt
Off, 0.5d, 1d, 3d
LIGHt
Low, Medium, High
B.LIGHt
Off, On, Auto
S.SAuEr
Off, 1min, 2min, 5min
(i-DT61PW)
Off, 5min, 10min, 30min
(i-DT61XW)
A.Off
Off, 5min, 10min, 30min
P.SAuER
ON, OFF
Off, 0.5min, 1min, 2min,
L.KEY
5min
End
\
EN-33
Options ( in
segment)
NO, yes
0.5d, 1d, 2d, 5d
2, 100
LOwW mMEd, HIGH
OFF, 0.5d, 1d, 3d
mMEd, HIGH, LOwW
OFF, ON, AUtO
OFF, 1, 2, 5 (i-
DT61PW)
OFF, 5, 10, 30 (i-
DT61XW)
OFF, 5, 10, 30
ON, OFF
OFF, 0.5,1, 2, 5