Controlling Stimulation Amplitude
2
3
1
8
3
7
22 22
110-0085 rev E.indd 22
2
5
4
6
9
7
Automatic Amplitude Adjustment
In the automatic amplitude adjustment mode the stimulation
amplitude automatically increases from 0mA to the amplitude
value set by the user. The amplitude increase can be stopped, or
stimulation can be turned off, at any time.
To use the automatic amplitude adjustment feature:
Use the slider bar ( 4 ) and up and down arrows ( 3 ) to set a
desired stimulation amplitude.
Press "Manual" ( 7 ). The text will change to "Auto" and
the button will be dark gray when automatic amplitude
adjustment is active.
Press the stimulation button ( 8 ) to turn on stimulation.
The stimulation amplitude will start increasing from 0mA
at a rate of 0.2mA per second, and the output stimulation
amplitude will be visible in the digital display ( 2 ) at the top
of the stimulation bar. A red bar on the stimulation bar will
represent the stimulation amplitude as it increases ( 9 ).
Note: The stimulation bar and up and down arrows are not accessible to
adjust stimulation.
To stop the increase in stimulation amplitude while still
delivering stimulation, press "Auto" ( 7 ).
To stop stimulation altogether, press the stimulation button
( 8 ).