Connecting to a Stimulator
If the Connect screen is opened after implanting a PNE lead
(from the PNE Lead implant screen), the CP will automatically
2
search for the nearest Trial Stimulator ( 1 ).
1
If the Connect screen is opened after implanting a Tined lead
(from the Define Thresholds screen), the "IPG Scan" or "EPG
Scan" button ( 3 ) should be selected to initiate a search for
3
Neurostimulator or Trial Stimulator respectively.
A scrolling search icon will appear when the CP is searching for
Stimulators ( 2 ).
The CP will display the nearest patient device ( 4 ). Check the
Device ID and the Patient ID to determine if this is the correct
device. New devices will not have Patient IDs.
Press "Connect" ( 5 ) to connect to the displayed device. The CP
4
5
will display a prompt while it connects to and retrieves data from
the Stimulator ( 6 ). After successful connection to the device,
8
the CP will automatically open the Patient Device screen.
Additional devices can be displayed by the CP if the first device
found is not the targeted Stimulator. Press the down arrow
next to "Show Additional Devices" ( 7 ). This will display up to 3
additional devices found in proximity of the CP.
Note: Hold the CP within 1 meter (~ 3 feet) of the targeted device to find the
device. If the desired device is not found, move closer and search again by
(
)
pressing "EPG Scan" or "IPG Scan"