Viewing the Status of a Trial Stimulator
E0
E1
E2
E3
44
110-0085 rev E.indd 44
1
2
Therapy History Tab ( 1 )
This tab shows the stimulation settings saved to the Stimulator
in the previous 8 stimulation programs. For each set of
stimulation settings, additional information is provided including
the time period the settings were used and notes entered during
the programming session. The usage information includes:
Daily Usage: Provides a percentage that reflects the daily use,
•
or on time, of stimulation.
Weekly Adjustments: Displays a numeric value of weekly
•
adjustments based on adjusting stimulation levels by Patient
Remote. Momentary adjustments will not be recorded.
Average Amplitude: Displays average stimulation amplitude
•
based on usage and adjustments.
Threshold Detail Tab ( 2 ) (Tined Lead only)
This tab summarizes the stimulation threshold data previously
captured. For each electrode, the tab shows the motor
response, sensory response, and threshold stimulation
amplitude recorded during the tined lead implantation
procedure or during a programming session. The color around
the threshold amplitude indicates the qualitative assessment of
the electrode placement (Red = "Poor", Yellow = "OK",
Green = "Good").
5/7/2020 11:53:37 AM