Setting Up a New Trial Stimulator

1
2
3
38
110-0085 rev E.indd 38
There are several steps to set up a new Trial Stimulator.
A. A prompt will appear when a new Stimulator is connected
if the trial type has not yet been assigned to the Stimulator.
Select the type of trial and the Trial Stimulator will be
configured appropriately.
B. Input the serial number of the patient's Remote Control
in the Remote Control ID field ( 1 ). This pairs the Remote
Control to the Stimulator, allowing the Stimulator to receive
commands from the Remote Control.
Note: The Stimulator can only communicate with the Remote Control that is
entered in this field.
C. Enter the patient identifier associated with this device into
the Patient ID field ( 2 ).
Note: When considering what patient ID to use, follow the applicable
guidelines from your organization regarding patient privacy.
D. Enter the date of the implant procedure in the Trial Start Date
field ( 3 ).
These fields, including the Patient ID and the Remote Control
serial number can be updated at any time as needed.