Foramen Needle Test Simulation
24 24
110-0085 rev E.indd 24
Pressing "Lead Placement" on the Home screen will present
a window for selecting which type of lead is being implanted.
Select the type of lead being implanted, and then the Foramen
Needle Placement screen will appear. This screen shows a
visual representation of the foramen needle ( 1 ) and allows
test stimulation of the foramen needle. To stimulate using the
foramen needle:
•
8
5
Note: The test stimulation cable is sterile and can be used in the sterile field.
Note: The stimulation ground cable is not sterile. Do not place the ground
6
electrode in the sterile field.
1
Note: The CP is not sterile and cannot be sterilized. When using the CP in a
sterile field, place it into a sterile bag.
9
4
7
2
3
Connect the foramen needle test stimulation cable to the CP
( 2 ) and to the foramen needle. Also connect the stimulation
ground cable to the CP ( 3 ) and to the patient (see the
lead implant manuals for more details on connecting to the
foramen needle and the patient).