Foramen Needle Test Simulation
26 26
110-0085 rev E.indd 26
8
5
6
1
9
4
•
7
2
3
•
Gray
(unassigned): Stimulation level is zero
•
Green
("good"): Stimulation level between 0.35 and 2 mA.
•
Placement is acceptable.
Yellow
("OK"): Stimulation level between 0.00 and 0.30 mA
•
or between 2.10 and 3 mA. Axonics recommends adjusting
the needle placement.
Red
("not recommended"): Stimulation level between 3
•
and 12.5 mA. Axonics strongly recommends adjusting the
needle placement.
Press ( 8 ) in the upper right to move on to place the lead.
For a PNE lead implant, this will say "PNE Lead Placement"
•
For a tined lead implant, this will say "Define Thresholds"
•
To exit back to the Home screen, press "Exit" ( 9 )