zero point potential, at this time you need to calibrate zero point potential (zero point potential
should be 0.00mV, you can select pH measuring parameter and view displayed zero point
potential value in measuring mode). For detailed zero point potential calibration, please refer to
the relative chapter about pH or pX measuring mode, usually when you enter into measuring
mode, press 'SETUP' key, select 'Calibrate zero point potential' and then press 'Enter'.
c. To protect and to use the meter better, before turn on the meter, please check the pH electrode
socket on backside of meter and make sure it is connected with measuring electrode or short
circuit plug, otherwise the high resistance apparatus of the meter will be damaged.
d. If meter is not used, please connect the short circuit plug to avoid damage of the meter.

4.3 Start Mode

The starting mode is shown as following chart, the upper left side of the screen
shows present system time. On the bottom is present measuring mode including measuring
parameter. On the right side is 'SETUP', 'VIEW' and 'MEAS' key. Press 'SETUP' on upper right
side, you can set 'Measuring Mode', 'Ion Mode', 'System Time', 'Manual Temperature' and
'System Setup' etc... Press 'VIEW' in the middle, you can view 'pH STD Data', 'pX STD Data',
'Conductive Measuring Parameter', 'TDS Factor', 'Dissolved Oxygen measuring parameter',
'View Stored Data'. On lower right side is 'MEAS' key, you can start measuring by pressing this
key.
4.4 Set Function in Start Mode
In this function, you can set 'Measuring Mode', 'Ion Mode', 'Time and Date', 'Manual
Temperature' and 'System Setup' etc... Under starting mode, pressing 'Setup', meter displays setup
menu which shown as follows:
Display chart of starting mode
59