Display chart for setting measuring mode
On left side, it is present selected measuring parameters, i.e. main measuring parameter chart. In
the middle is measuring mode including Continuous Mode, Timed Reading measuring mode and
Balance Reading measuring mode. When selected parameter is ion concentration, measuring
mode would be changed to concentration measuring mode respectively which includes Direct
Reading Mode, Known Addition mode, Sample Addition mode and GRAN Method mode, etc...
On the right side of the screen is press key, press 'MEAS' to start measurement. In case you want
to start the same measurement next time, you can press 'ENTER' key, meter will save all present
setting automatically and then return to Start Mode. Or you can press 'CANCEL' to exit and
return to Start Mode.
On the bottom of screen, meter displays present measuring result which is shown as follows:
The actual display will vary from the value of selected main measuring parameter. Each
measuring part has one individual measuring window which displays the name of main measuring
parameter, measuring value of main parameter and measuring value of auxiliary parameter. There
is a '>' button on the left side of the window, you can select directly the other main measuring
parameter under present part. For example, press '>' button beside conductivity window, meter
will switch main measuring parameter to TDS measuring, press again, switch to salinity
measuring, press once more, switch back to conductivity measuring (note: this method do not
suitable for ion measuring part).
There is a "▼"or "▲" key on the right side of each measuring window, press this key to switch
the displaying position of main measuring parameter ( note: this method do not suitable for
individual main measuring parameter).
4.4.1.1 Select Measuring Parameter
Press 'Parameters' displayed on the left side of screen, meter shows 'Select Parameters' window
which shown as follows:
Measuring
display
conductivity, pH, DO measuring parameter.
chart
when
select
61