allows, whether you agree to cover or repeat the storage. For example, the permitted stored pH
measuring data is 200 sets, if you want to store number 201 data, in case automatic delete stored
data function is opened, meter will store the number 201 data in first position, that means meter
will store data from the beginning, otherwise meter will give up storing current data. Please pay
special attention on it.
GLP function means meter has two modes, i.e. either select supporting GLP or do not select
supporting GLP. There will be different pattern in search and print stored data under these two
modes.
'Balance Reading' measuring condition is in correspondence to the Balance Reading condition of
each measuring parameter, displayed pH balance condition is 0.01pH. When change amount of pH
measuring value is less than 0.010pH, the measuring is valid. This function is mainly for 'Balance
Reading' measuring mode. When user selects 'Balance Reading' measuring mode to measure and
if all measurement within set balance time are accord with balance condition, the current
measurement finishes.
Balance time is only valid in 'Balance Reading' measuring mode and the unit is second (s), range
is 1 ~ 200 seconds.
4.4.6 Set Measuring Parameter
Select 'Set Measuring Para.', meter jumps out a selectable menu which displayed as follows:
User can enter into relevant parameter set function part according to practical need and press
'CANCEL' to quit.
4.5 View Function
For easy use, meter allows user to view present measuring parameter at random including
previous calibrating data and present using parameter, allows user to edit measuring parameter
directly. Meter also allows view stored data.
In start mode, press 'View' key, select relevant item to view, edit measuring parameter and view
stored data which displayed as follows:
Displayed
chart
of
Parameter'
'Set
Measuring
68