Descargar Imprimir esta página

J.P. SELECTA MP-2006 Manual De Instrucciones página 89

Publicidad

Idiomas disponibles

Idiomas disponibles

When all setting finish, you can start measuring of known addition concentration mode which is
displayed as follows. In which on upper side of the screen is the present measuring data, at the
bottom is the records of measuring result including potential, temperature before and after adding.
On right side is 'ENTER', 'END' and 'SETUP' key.
Press 'SETUP' key, you can set manual temperature, select concentration unit, set mV display
resolution, pX display resolution which is already been calibrated potential zero point.
Clean the relevant ion select electrode and put in measured solution, meter displays current
potential, temperature value and relevant pX value.
When display stable, press 'ENTER' key, meter stores present potential and displays 'Add
Standard Solution', you can add standard solution according to set volume, when display stable
again, press 'ENTER', meter displays 'Measuring Finish!' and calculate the concentration value
of present sample which is displayed as follows:
At this time you can store present measuring result or print the result.
Similarly, if you want to select the concentration unit of sample's concentration, you can also
operate according to following procedure: press once at display area of 'sample Cs', the
concentration unit will switch for one time.
Known
addition
measuring mode
Display chart when known adding
concentration
measuring
finishes
concentration
mode
86

Publicidad

loading

Productos relacionados para J.P. SELECTA MP-2006

Este manual también es adecuado para:

4120700