Display chart of 'View' in start mode
4.5.1 View pH STD Data
Under this function, user could view current pH calibration data. In start mode, press 'View' key,
then press 'Enter' to view 'pH STD Data'. Displayed as follows:
Display chart of 'View pH STD Data'
On top of the screen is last calibration data including calibration time, calibration point,
calibration data and current slope in percentage in correspondence to each STD. At the bottom is
the condition of STD buffer group: on right side is "<=>"(move) key, 'PRINT' key, 'STD' key
and 'SETUP' key..
If last calibration is multi-point calibration, user can press "<=>"(move) key to show detailed
data of each standard solution point.
Here, user could calibrate electrode slope (please refer to calibration chapter for detailed
calibration procedure) directly by pressing 'STD'.
Press "SETUP' key, user could set new STD buffer group.
Press 'X'(quit) key to exit 'View pH STD Data'.
If user wants to print present calibration data, you could connect the connecting line (attached
with meter) with printer TP-16 (should be bought separately) and press 'PRINT' to print out
calibration data.
69