Please press 'PRINT' to print out calibration data when the meter is connected with model TP-16
printer.
4.5.3.1 Setup 'Cell Constant'
Generally, each conductive electrode from our company has relevant conductive constant, you
only need to set constant value on electrode once before using. After using a period of time, if you
suspect the electrode constant is not precise, you can calibrate with conductive standard solution
again. Press 'STD' key, you can calibrate electrode slope (please refer to chapter Calibration for
detailed procedure).
So there are two ways to set present cell constant value: one is to re-calibrate with conductive
standard solution, as soon as calibration finishes, meter will calculate cell constant automatically.
Another ways is to set cell constant value directly.
You can only select one method. Presume you get cell constant value through calibration the
previous time and now you want to input cell constant directly, meter will delete previous
calibration data, so please take care.
Press 'SETUP' key beside 'Cell Constant', there will be an input window, please input new cell
constant then press 'ENTER'. Displayed as follows:
Press 'X'(quit) to exit 'Set Cell Constant' mode.
Display of View 'conductive measuring
parameter'
Display chart of 'Set Cell Constant'
72