3.
Application
What you should know when connecting the bipolar instrument
3.1.
Preparation
HF cable
Active electrodes
HF unit
8
Avoid monitor display interference:
•
Do not run HF cable directly parallel to camera cables.
•
Do not lay HF cable in loops.
•
Never place HF cable onto the skin of the patient.
Connect instrument for coagulation and switch on
electrosurgical generator.
Grasp HF cable only by the connector when plugging in and
unplugging.
WARNING
Risk of unintended burns.
Never use worn out or faulty active
Dispose of worn out or faulty active
Pay attention to the electrosurgical generator
instruction manual and to general information
on electrosurgical procedures.
Technical data, page 7
WARNING
Improper use of HF current can cause endogenous and
exogenous burns as well as explosions.
Carry out electrosurgical procedures only with
insufflation of incombustible gases (CO
electrodes.
electrodes - never repair!
).
2
MN030-667-S2 EN