Bowa ErgoLAP Manual Del Usuario página 30

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 43
Adjusting the
output power of the
electrosurgical
generator
Safety measures
10
WARNING
Risk of unwanted burns.
 Before the procedure, try out suitable approximate
values extracorporally.
 Set the output power of the electrosurgical
generator only to the value absolutely required for
the procedure.
 Technical data, page 7
 Before increasing the output power due to
insufficient coagulation, without fail check:
Perfect contact of all HF cables and connectors.
Foot switch function.
Insulation of the HF cable, active electrodes, trocar
sleeve or shaft,
Cleanliness and wear of the distal end of the active
electrodes.
 Switch on the HF current only when the active
electrodes are in contact with the tissue to be
coagulated (don't use Auto Start).
 Active HF instruments must not be placed upon
the pacient: this may lead to burns.
 Using plastic trocars increases the electrical
safety.
MN030-667-S2 EN

Publicidad

Tabla de contenido
loading

Tabla de contenido