3.6.
Repair
Do not repair or service defective devices.
Discard or replace defective devices.
4.
Preparation
Ensure that insulated parts do not come in contact with hard, pointed or heavy
objects during preparation, as such objects may damage the insulation and render
the product unsuitable.
Preparations cycles:
Handle, outer and inner shaft tube: 200 cycles
Jaw: 75 cycles
1. Soaking
Soak the instrument immediately after use, or no later than 2 hours after use. Use
only aldehyde-free disinfectants (recommended: Gigasept Instru AF) designed for the
disinfection of instruments (e.g. DGHM or FDA approval, or CE mark).
Place the instrument in an ultrasonic bath for at least 5 minutes.
Use a cleaning solution to soften coagulation residues (encrustations) at the jaws and
remove them with a soft cleaning cloth.
A synthetic feece or plastic cleaning brush can be used if necessary for the
preliminary removal of residues.
Do not pre-clean HF cables in the ultrasonic bath.
2. Disasembly
Disasembly, page 14
MN030-667-S2 EN
15