The last three stages of this cleaning process (rinsing, drying and lubrication) may be performed by automatic devices available on
FR
the dental market, following the instructions of use of these devices.
MICRO-MEGA
recommends using X-Cidol
®
EN
3.3 Automatic cleaning prior to sterilization
(Follow instructions after every use.)
DE
In order to eliminate all handling, after each use on a patient place the contra-angle in a device of the product line X-Cid
These devices clean, disinfect and lubricate your handpieces and contra-angles without previous dismantling and guarantee a rapid
IT
bactericide, fungicide and virucide effect, ensuring the destruction of microbial agents such as HIV, HBV and Tuberculosis.
The regular use of a device of the product line X-Cid
ES
3.4 Maintenance (Lubrication)
(Follow instructions after every use.)
RU
•
Remove the root canal instrument.
•
Introduce the "500l" spray into the rear end of the contra-angle.
•
Press the spray a number of times lasting several seconds until the liquid coming out of the contra-angle is clean.
20
2 as pre-disinfectant product.
®
®
significantly extends the life span of your handpieces and contra-angles.
•
Make the contra-angle turn in order to evacuate excess oil before use.
•
During long treatments, repeat the lubrication process every ten minutes.
3.5 Sterilization
(Follow instructions before every use.)
•
MICRO-MEGA
®
handpieces and contra-angles can be sterilized (wrapped in a sterilization bag) in an autoclave at 135°C (275°F) to
138°C (280.4°F) and 2.2 bar during 18 minutes.
®
.
•
Following manual or automatic preparation for sterilization and lubrication, place the contra-angle in a hermetic sterilization bag
(wrap the contra-angle separately in a retractable sterilization bag):
- Individual packaging: A standard packaging material may be used. Make sure that the sterilization bag is big enough for the contra-
angle so that using force to close the bag is not necessary. The sterilization bag must be of appropriate size for the contra-angle
and not be distended.
- Batch processing: The contra-angle may be placed in an instrument box provided to this effect or in a tray generally used for
sterilization. Wrap up the trays using appropriate methods.
•
Perform sterilization in a sterilizer compliant with the standard EN 13060. Contact corrosion due to humidity may be possible.
•
In case you want to sterilize several handpieces or contra-angles in one autoclave cycle, make sure that the autoclave's maximum
load is not exceeded (cf. the autoclave's instructions of use).