English
Osstell ISQ Preparation and Set-up
Batteries
The internal battery is rechargeable and should be charged for at least 3 hours, before its first use.
Setting the Time
Prior to the use of the instrument, the date and time should be adjusted (see set-up). The date and
time is stored when each implant measurement is performed.
Test Peg
The Test Peg included in the Osstell ISQ package may be used for testing and learning how to use
of the system. The Test Peg may be used as follows: Place the Test Peg on a table or hold it by hand.
Turn on the instrument and hold the measurement probe (See Figure 2) close to the top of the Test
Peg, until the instrument beeps and displays the ISQ value. The point to aim the probe tip at is
marked with red paint on the Test Peg.
Note: Do not remove the SmartPeg from the block!
Osstell ISQ Operation
Operation
To turn the instrument on, press any key. To turn it off, enter the setup menu by pressing the centre
key and press the arrow up or down keys until "Power off" is marked. Then press the centre key
again. In measurement mode, the instrument will power down automatically after 2-30 minutes
(adjustable) of inactivity. To return to the measurement mode from the menu mode, press the left
arrow key.
The instrument is to be used with the probe connected to the instrument via its cable (Figure 1).
The probe with its cable can be autoclaved.
Note: Do not twist the connector when the probe is connected! To remove it from the instrument,
pull gently holding around the connector.
Connecting SmartPeg
SmartPeg should be handled carefully, as damages to the SmartPeg may affect the measurement
result. Connect the SmartPeg Mount to a SmartPeg (Figure 3). The SmartPeg is magnetic, and the
Mount will hold the SmartPeg as it is carried to the implant. Screw the SmartPeg onto the implant
or abutment. Use approximately 4-6 Ncm of torque. Do not over tighten, to avoid destroying the
SmartPeg threads. The SmartPeg is disposable, and may be used for 10-20 attachments during one
patient's session.
Note: The SmartPeg is a single-use device and should not be re-used due to the risk of false readings
or even no reading at all. Re-use can also result in a broken SmartPeg or damaged SmartPeg threads.
Performing a Measurement
Attach a SmartPeg to the implant or abutment. Hold the measurement probe close to the top of the
SmartPeg without touching it (Figure 2). Do not press any keys. When the instrument senses the
SmartPeg and the measurement is ok, it will emit an audible sound. If two such sounds are heard
21