Calibrating the Device
1
2
14
3
If you own multiple devices make sure that the serial number on the front and back
of the calibration plate matches the serial number of your i1Pro 3 device. Refer to the
user guide of your measurement software for additional information on the calibration
process.
1. In the center of the calibration plate is a slider to protect the white calibration tile.
Open this protective slider until it clicks into place to access the white reference
ceramic tile on this calibration plate.
2. Place the i1Pro 3 measurement device on the calibration plate. The device must
seat firmly on the calibration plate.
3. Start the calibration mode in the software, and press the measurement button on
the device or in the software. Status indicator light switches off during calibration.
Once the device is successfully calibrated the status indicator light pulsates white.
If the calibration fails the status indicator light pulsates red and then returns to
solid white.
4. When the calibration is finished you should close the protective slider for the white
reference ceramic tile.
A clean white reference ceramic tile is essential for providing accuracy of
your measurements. The protection cover should always be closed when
the white reference ceramic tile is not in use. If necessary you can clean the
white reference ceramic tile with isopropyl alcohol and a soft clean cloth.