Power Module Cleaning; Manual Cleaning - Stryker SYSTEM G Manual De Instrucciones

Ocultar thumbs Ver también para SYSTEM G:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 13
7300-001-700 Rev-B
CAUTIONS:
▪ DO NOT use solvents, lubricants, or other chemicals,
unless otherwise specified.
▪ DO NOT use automated washers or ultrasonic cleaning
equipment.
▪ DO NOT immerse or soak any equipment in liquid.
Handpieces and attachments may be placed briefly
under running water to ensure thorough wetting and
contact with liquid while actuating moving parts.
▪ DO NOT allow moisture or liquid to enter electrical
plugs, receptacles, or openings. Moisture or liquid may
cause corrosion and damage the equipment.
▪ DO NOT use pipe cleaners or cotton swabs to clean
lumens or connector receptacles.
▪ ALWAYS make sure the detergent solution is
completely rinsed off before drying the equipment.
▪ Use of compressed air is only recommended for drying
of equipment.

Power Module Cleaning

CAUTION: DO NOT rinse, spray, or wipe the power
module with a saturated cloth.
NOTE: Keep the power module as clean as possible, as it
cannot be sterilized.
1. Wipe all debris from external surfaces of the
handpiece door.
2. Remove the power module from the handpiece. Follow
the steps in the instructions for use manual supplied
with the handpiece.
3. Wipe the power module with a clean, dry cloth.
4. After cleaning, inspect the power module. See the
Inspection section.

Manual Cleaning

CAUTION: DO NOT use these instructions to clean power
modules. See the Power Module Cleaning section.
1. Remove the lid from the sterilization case.
2. Remove all soiled medical devices from the
sterilization case.
3. Remove all gross soil from the equipment using
absorbent wipes or a soft, lint-free cloth moistened
with the prepared detergent solution. Devices may be
placed briefly under running water or flushed with a
water-filled syringe to assist in the removal of gross
soil.
www.stryker.com
4. Thoroughly clean the equipment. See Appendix A:
Special Cleaning Considerations for critical areas and
brush sizes.
▪ Use suitable brushes to clean all surfaces. Pay
particular attention to rough surfaces, crevices, and
difficult-to-reach areas where soil may be shielded
from brushing.
▪ Thoroughly brush the equipment for a minimum of
two minutes.
▪ Open the door on the bottom of the handpiece and
use suitable, non-abrasive brushes to clean the
interior of the power module compartment.
▪ Use soft brushes of appropriate diameters to clean
the entire length of each lumen. For dead-ended
lumens, use light pressure and do not force the
brush any farther after you feel resistance.
▪ Use a syringe filled with detergent solution to flush
difficult-to-reach areas.
▪ Actuate all moving parts to clean hidden surfaces.
▪ When removing debris from a reciprocating saw,
rotate the blade collar to access all surfaces.
5. Thoroughly rinse the equipment with warm, running
water until all traces of detergent solution are removed:
▪ For equipment with lumens, hold the equipment
on an incline, distal end pointing up, and rinse the
lumen.
EN
5

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

7302-120-0007302-130-000

Tabla de contenido