OBJ_DOKU-193153-006.fm Seite 32 Mittwoch, 25. März 2020 4:12 16
8.7
Endoscope selection
The HF leakage current at the ocular of the endoscope may
cause thermal injury to the user. In order to prevent this, please
use only endoscopes designed for use with electrosurgical devic-
es.
The working channel of the endoscope must have a diameter of
at least 2.5 mm.
8.8
Using the product
The water jet and APC are alternately activated via different
footswitches.
The water jet is activated via the ERBEJET 2 footswitch, the APC
via the COAG pedal of the VIO footswitch.
9
Disposal
Dispose of the product, packaging material and accessories (if
available) in accordance with the national guidelines and legis-
lation applicable in each case.