ELAN EC motor with right/left rotation (without irrigation unit/with irrigation
unit)
7.1

Fuse replacement

Risk of fatal injury from electric shock!
Unplug the device before changing the fuses!
DANGER
Specified fuses: 2 x IEC 127 - T4.0 A, Breaking capacity L (L=40 A at
250 V/50 Hz).
Use a small screwdriver to release the clip on the fuse holder 21.
Remove fuse holder 21.
Replace both fuses.
Reinsert fuse holder 21 so that it audibly snaps into place.
Note
If the fuses burn out frequently, the device is faulty and should be
repaired, see Technical Service.
14
®
8.

Technical Service

Risk of injury and/or malfunction!
Do not modify the product.
WARNING
For service and repairs, please contact your national B. Braun/
Aesculapp agency.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.