Stacking Of Units; Presetting The Correct Voltage; Connecting The Power Supply - Braun Aesculap Instrucciones De Manejo/Descripción Técnica

Ocultar thumbs Ver también para Aesculap:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 52
Aesculap
ELAN EC motor with right/left rotation (without irrigation unit/with irrigation
unit)
3.2

Stacking of units

Note
The device can be assembled on a height-adjustable Aesculap mobile stand
GA415 or on an Aesculap mobile stand for surgical devices GD416M.
Note
Do not stack other devices on top of the device.
Assembly/disassembly on mobile stands according to the instructions
for use of the mobile stand.
Make certain the device is set up on a sufficiently stable support (table,
equipment cart).
Ensure that the console is set up in a stable position.
3.3

Presetting the correct voltage

Risk of fatal injury from electric shock!
Pull the mains plug before unlocking the fuse
holder; otherwise you could risk touching com-
WARNING
ponents that may be live with potentially lethal
mains voltage.
The mains voltage must correspond to the voltage indicated on the type
plate at the back of the unit.
The correct voltage for the device is indicated in the window of the fuse
holder located at the back of the device. The standard voltage for the unit
in Europe is: 230 V.
If the registered voltage does not match the mains voltage, the setting
must be changed at the back of the control unit as follows:
Release clip B using a suitable screwdriver A and remove the fuse
holder 21.
Remove the voltage switching insert 20 underneath it.
Reinsert the voltage switching insert 20 so that the desired voltage is
legible.
6
®
Insert the fuse holder 21 until it audibly engages.
20
C
Fig. 1
Legend
20 Voltage switching insert
21 Fuse holder
22 Device connector
A Screwdriver
B Clip
C Voltage switching insert imprint
Imprint
Voltage range
110
100 V~–120 V~ (+/-10 %)
230
220 V~–240 V~ (+/-10 %)
3.4

Connecting the power supply

Risk of fatal injury from electric shock!
Connect the product only to a grounded power
DANGER
supply.
The mains voltage must correspond to the voltage indicated on the type
plate at the back of the unit.
Plug in the power cord at device connector 22.
Connect power supply by inserting into a wall socket.
22
A