WARNINGS & CAUTIONS
Failure to comply with precautions and warnings described in this User Manual may lead to exposure to dangerous optical radiation sources. Please comply with all safety instructions and warnings.
CAUTION:
- Read these instructions carefully prior to using your Gemini 810 + 980 diode laser.
- Ensure that all users are properly trained prior to use. Consult your distributor for training recommendations.
- Do not modify this equipment without authorization of the manufacturer.
- Always wind the fiber optic cable in a clockwise manner around the fiber wrap to avoid fiber breakage.
- Do not use in the presence of combustible or combustion-supporting gases.
- Always test activate the device outside the mouth before using on a patient.
- This unit has been designed and tested to meet the requirements of electromagnetic, electrostatic, and radio frequency interference standards. However, the possibility of electromagnetic
or other interference may still exist. Relocating the device may help to eliminate the interference.
- Medical electrical equipment needs special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information
provided in the Electromagnetic Environment Guidance section of this manual.
- Periodically inspect laser eyewear for pitting and cracking.
Safety is paramount when using any energy-based surgical instrument, and your office should implement a safety program for the Gemini 810 + 980 diode laser. If your office
does not already have a safety officer, one should be appointed to be responsible for understanding proper use, safe operation, and maintenance of the Gemini laser system.
Their duties should include training office personnel in all aspects of system safety and management of the Gemini laser and all accessories.
WARNING:
- Visible and Invisible Laser Radiation – Avoid eye or skin exposure to direct or scattered radiation. Class IV Laser Product.
- Laser Safety Eye Protection MUST BE WORN by the Operator, Patient, Assistant, and anyone present when the laser is activated. Eye Protection must conform to Specification DIN EN207
Annex II of the Directive 89/686/EEC with wavelength protection of 810 nm- 980 nm, and ±10 nm of OD 5+ such as NoIR Laser company filter model CYN.
- Never direct or point the beam at a person's eyes.
- Do not look directly into the beam or at specular reflections.
- Do not aim the laser at metallic or reflective surfaces, such as surgical instruments or dental mirrors. If aimed directly at these surfaces the laser beam will reflect and create a
potential hazard.
- Never operate the laser without a fiber tip attached.
- Laser aperture at the end of the handpiece.
- Laser aperture warning label affixed to system handpiece.
- Always place the system into STANDBY mode when leaving the Gemini 810 + 980 diode laser unattented for few minutes or between patients.
- Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
- Do not open unit housing at any time. Danger from optical radiation may exist.
- The use of accessories, other than those specified, except those supplied or sold by Ultradent Products, Inc., as replacement parts for internal or external components, may result in
increased EMMISSIONS or decreased IMMUNITY of the Gemini 810 + 980 diode laser.

WHAT IS IN THE BOX

The Gemini 810 + 980 diode laser includes the following:
Laser Unit
Activation Pedal with 2 AA Batteries
Disposable Tips (10)

UNPACKING INSTRUCTIONS

A manufacturer or dealer representative can provide assistance when you are ready to remove the Laser from its shipping container. Please do not attempt to unpack the Gemini 810 + 980 diode laser
and install the system without reading this manual first. If you are unsure about any aspect of the assembly, call your customer service representative or dealer for assistance.
SHIPPING CONTAINER INFORMATION
The shipping container you received with your Gemini 810 + 980 diode laser was specially designed to safely transport the device. In the unlikely event that you need to return the laser for service or
repair, please retain the original shipping container.
2
Fiber Delivery System
*Actual Power supply may vary from image
*Actual eyewear may vary from image
*DC Power Supply
*Protective Eyewear (3)
Laser Warning Sign
Warranty Information
USB Cable
User Manual
NOTE: The laser ships with the lithium ion battery and fiber
delivery system already installed
NOTE: Use proper care when transporting the unit
WARNING: No modification of this equipment is allowed